Liquid chromatography is the separation technique behind almost every purity figure printed on a peptide certificate. A liquid carries the sample through a packed column, the components of the sample travel at different speeds, and a detector records each one as it emerges. To judge a purity result rather than simply trust it, you need to know which kind of liquid chromatography produced it.
This guide defines liquid chromatography, compares its six main modes, explains why peptides are usually run in one of them, and walks through the modules of a high-performance liquid chromatography (HPLC) system and the checks that link it to a certificate.
The short answer
- Liquid chromatography is a separation method. Components are distributed between a stationary phase and a moving liquid mobile phase, following IUPAC’s definition of chromatography.
- Separation comes from differential retention. Each component interacts with the stationary phase to a different degree, so each leaves the column at its own retention time.
- There are six common modes: reversed-phase, normal-phase, hydrophilic-interaction (HILIC), ion-exchange, size-exclusion and affinity, each separating by a different property.
- Peptide purity is usually measured by reversed-phase HPLC, with a water-to-organic gradient and an acidic ion-pairing reagent such as trifluoroacetic acid (TFA).
- An HPLC system is a chain of modules: solvents, pump, autosampler, temperature-controlled column, detector and data system, each of which affects the final number.
- Peptides are detected by the peptide bond, which absorbs ultraviolet light near 214 nm.
- A purity figure is only as good as its method, which is why pharmacopoeial system-suitability tests and ICH validation exist.
Liquid chromatography (LC)
LC · HPLC (high-performance liquid chromatography) · column chromatography
- What it is
- A separation technique in which a liquid mobile phase carries a sample through a stationary phase
- Defined by
- IUPAC nomenclature for chromatography (Recommendations 1993)
- Origin
- Mikhail Tswett, 1906: plant pigments separated on a column of calcium carbonate
- Main modes
- Reversed-phase, normal-phase, HILIC, ion-exchange, size-exclusion, affinity
- Usual mode for peptide purity
- Reversed-phase HPLC with an acidic ion-pairing mobile phase
- Usual peptide detection
- UV absorbance near 214 nm, where the peptide bond absorbs
- Instrument modules
- Solvent delivery, sample introduction, column and oven, detector, data system
Where the evidence stands
- Standardised method
- Validated in labs
- Reported on COAs
- Comparable between labs
Mature and standardised; two purity figures compare directly only when their methods are stated and similar.
What liquid chromatography is
Liquid chromatography (LC) is a family of separation techniques in which a liquid, the mobile phase, flows through a stationary phase, usually fine particles packed into a column. IUPAC’s definition of chromatography is deliberately general: the components to be separated are distributed between two phases, one stationary and one moving in a definite direction.
The separation itself comes from differential retention. A molecule that interacts strongly with the stationary phase leaves the column later; one that interacts weakly is swept through faster. The time from sample introduction to the top of a component’s peak is its retention time. Pushing components out with the mobile phase is elution; changing the mobile-phase composition during the run is gradient elution.
The detector’s record of this is a chromatogram: signal against time, one peak per separated component. Reading that trace is covered in our guide to reading an HPLC chromatogram; this page is about how the separation is produced.
From a chalk column to HPLC
The method began with plant pigments. In a 1906 paper in the Berichte der Deutschen Botanischen Gesellschaft, the botanist Mikhail Tswett described packing a fine dry powder such as precipitated calcium carbonate into a glass tube, washing a leaf-pigment extract through it, and watching the pigments resolve into coloured bands. He called the banded column a chromatogram; an English translation appeared in 1967.
The theory that made chromatography quantitative came from A. J. P. Martin and R. L. M. Synge. Their 1941 Biochemical Journal paper described a chromatogram employing two liquid phases, the basis of partition chromatography; they shared the 1952 Nobel Prize in Chemistry “for their invention of partition chromatography”. In 1956, van Deemter, Zuiderweg and Klinkenberg analysed why real peaks spread out, attributing non-ideal band broadening to longitudinal diffusion and resistance to mass transfer. Our HPLC vs UPLC comparison picks up how particle size follows from that theory.
Other principles followed: Porath and Flodin’s gel filtration in 1959, the ancestor of size-exclusion chromatography, and Cuatrecasas, Wilchek and Anfinsen’s affinity chromatography in 1968. In 1967 Horváth, Preiss and Lipsky reported “fast liquid chromatography”, separating nucleotides on pellicular ion exchangers, an early step toward the instrumented technique now called HPLC.
Canada has a genuine place in the peptide chapter of this history. In 1986, Robert Hodges’s group at the University of Alberta in Edmonton, part of the Medical Research Council of Canada Group in Protein Structure and Function, measured retention coefficients for individual amino acid residues in model synthetic peptides, so that reversed-phase retention could be predicted from sequence. Hodges and Colin Mant later continued it at the University of Colorado. For more Canadian peptide science, see Canada’s century of peptide discovery.
The main modes of liquid chromatography
Every LC mode uses the same plumbing; what changes is the chemistry of the two phases, and so which property drives the separation.
| Mode | Separates by | Stationary phase (typical) | Role in peptide work |
|---|---|---|---|
| Reversed-phase (RP) | Hydrophobicity | Silica with bonded alkyl chains; mobile phase more polar than the column | The default for purity and for purification |
| Normal-phase (NP) | Polarity | Polar surface, such as bare silica; mobile phase less polar than the column | Rare for peptides; mainly small, organic-soluble molecules |
| HILIC | Hydrophilicity | Hydrophilic packing eluted with a mostly organic mobile phase | An orthogonal check: elution order is typically the reverse of RP |
| Ion-exchange (IEX) | Net charge | Charged groups that bind oppositely charged analytes | Resolves charge variants that RP can miss |
| Size-exclusion (SEC) | Molecular size | Porous particles that small molecules enter and large ones bypass | Detects aggregates and oligomers; desalting |
| Affinity | Specific binding | An immobilised ligand that binds the target selectively | Purifying proteins and binding partners rather than measuring purity |
Reversed-phase and normal-phase
The names are defined by polarity. In IUPAC’s terms, normal-phase chromatography is an elution procedure in which the stationary phase is more polar than the mobile phase, and reversed-phase chromatography is the opposite: the mobile phase is significantly more polar than the stationary phase, for example silica carrying chemically bonded alkyl chains. IUPAC also calls “reverse phase” “an incorrect expression to be avoided”.
HILIC
Andrew Alpert introduced hydrophilic-interaction chromatography in 1990 as a variant of normal-phase chromatography. A hydrophilic column eluted with a mostly organic mobile phase retains more hydrophilic solutes more strongly, so elution order is typically the reverse of reversed-phase, and its selectivity complements other modes; Alpert applied it to peptides, including phosphorylated peptides.
Ion-exchange, size-exclusion and affinity
Ion-exchange separates by charge, useful for impurities that differ from the target by a charged group but hardly in hydrophobicity. Size-exclusion separates by size, and the Hodges group’s methods review describes using it to assess a peptide’s oligomeric state. Affinity chromatography is a purification tool far more often than a purity test. For identity, LC is coupled to a mass spectrometer, a combination covered in LC-MS explained.
Why peptides are usually run in reversed-phase
Reversed-phase HPLC is the workhorse of peptide analysis because it can resolve sequences that differ by a single amino acid, and because its retention behaviour is well understood: it can even be predicted from sequence, as the Alberta work showed. A typical run ramps from a largely aqueous mobile phase toward an organic solvent such as acetonitrile, and peptides come off in order of increasing hydrophobicity.
The acid in the mobile phase matters as much as the column. Trifluoroacetic acid is the usual anionic ion-pairing reagent: it pairs with a peptide’s positively charged groups and sharpens peaks. A 2005 Hodges-group study compared TFA with more hydrophobic perfluorinated acids on ten-residue peptides carrying +1, +3 or +5 net charge: separation improved as the reagent became more hydrophobic and, to a lesser degree, more concentrated, and more highly charged peptides responded more.
Temperature is the other lever. Testing peptides between 5 and 80 °C, Chen, Mant and Hodges found that amphipathic helical peptides responded more strongly than random-coil peptides, attributed to helices unfolding during the run, and that temperature altered selectivity between peptides differing by one amino acid.
One method is not the whole picture. An impurity that co-elutes with the main peak is counted as product. Orthogonal methods such as HILIC, ion-exchange or size-exclusion exist to catch those cases; the Hodges group has compared reversed-phase with mixed-mode HILIC/cation-exchange separations of closely related peptides.
How reversed-phase purity is calculated and reported is covered in our HPLC peptide testing guide, and what a purity percentage does and does not mean is explained in peptide purity explained.
Our research peptides are lyophilized laboratory reagents, priced in CAD and shipped from inside Canada.
Browse research peptidesHow an HPLC system works
An HPLC system, often called a liquid chromatography machine, is a chain of modules, each controlling one variable. Walking through the HPLC instrumentation module by module shows where a purity result can go wrong.
Solvent reservoirs and degassing
Bottles hold the mobile-phase components, typically acidified water and an acidified organic solvent for peptide work. Dissolved gas is removed first, because bubbles disturb flow and detection.
Pump and gradient
The pump delivers a steady flow and blends the solvents into the programmed gradient. If the organic fraction arrives late or unevenly, retention times shift and close peaks can merge.
Autosampler and sample introduction
The autosampler draws a measured volume of sample into a loop, and a switching valve places that loop in the flowing mobile phase so the sample is swept onto the column. Variation in the volume loaded shows up as variation in peak areas, and carry-over from the previous sample is checked with blank runs.
Column and column oven
The stationary phase follows from the mode, particle size sets peak sharpness, and the oven holds a temperature that can change selectivity.
Detector
For peptides the detector is usually a UV absorbance detector set at a low wavelength. Kuipers and Gruppen measured the molar extinction coefficient of the peptide bond at 214 nm as 923 M−1 cm−1. Some residues absorb far more strongly: tryptophan roughly 30 times more than a peptide bond, and phenylalanine, tyrosine and histidine roughly six times more. So UV area percentage is a practical measure of purity, not an exact molar ratio.
Data system
The software draws the chromatogram and integrates each peak. Purity by area normalisation is the main peak’s area divided by the total integrated area, so where the baseline is drawn and which small peaks are counted are part of the method.
How does HPLC work, in one run
- The column is equilibrated in the starting mobile phase.
- The autosampler loads the sample and the gradient starts.
- Components leave the column in order of retention and are recorded as peaks.
- The data system integrates the peaks; the column is washed for the next sample.
System suitability and method validation
Because every module is a variable, regulated laboratories run a system suitability test before samples to confirm the complete system meets defined criteria that day. Typical checks are resolution between critical peaks, peak symmetry and the repeatability of replicate runs.
In the United States Pharmacopeia these requirements sit in General Chapter <621> Chromatography. A harmonized version, coordinated with the European Pharmacopoeia, became official on 1 December 2022. That revision added a definition of system sensitivity, a general peak-symmetry requirement of 0.8 to 1.8, and changed how signal-to-noise is measured.
System suitability checks the instrument on the day. Method validation checks that the procedure itself measures what it claims to. The ICH Q2(R2) guideline, in effect in the European Union since 14 June 2024, sets out how procedures for assay, purity and impurities, and identity are validated. Our guide to analytical method validation explains what those characteristics mean for a purity number.
From instrument to certificate
A certificate of analysis compresses a whole run into a few lines. A useful HPLC section should answer most of these questions:
- Which mode and column? Reversed-phase is expected for peptides.
- Which mobile phase and gradient? The acid used and the gradient range affect which impurities are resolved.
- Which wavelength? Near 214 nm detects the peptide backbone.
- Is the chromatogram shown? A trace shows whether small peaks were integrated.
- Is identity confirmed separately? Retention time alone does not prove identity; mass spectrometry does that job, as described in reading a peptide mass spectrum.
- Does the batch number match? A report is evidence only for the batch it names; see lot numbers and batch traceability.
Our peptide COA guide covers the rest of a certificate, and for multi-component products, how peptide blends are tested explains why a single purity figure can mislead.
What’s established and what isn’t
Reasonably well established
- Liquid chromatography separates components by their differential distribution between a stationary and a moving liquid phase.
- Reversed-phase HPLC with an acidic ion-pairing reagent is the standard method for peptide purity.
- Ion-pairing reagent and column temperature change peptide retention and selectivity.
- The peptide bond is the main UV chromophore at 214 nm; aromatic residues absorb more strongly.
- Pharmacopoeial system-suitability criteria and ICH validation define when a chromatographic result is fit for purpose.
Open questions
- Whether a single reversed-phase method has resolved every impurity in a given batch; co-elution can only be ruled out with orthogonal methods.
- How closely UV area percentage tracks true molar purity for a particular peptide, since impurities absorb differently.
- Whether purity figures from two laboratories using different columns, gradients and wavelengths are directly comparable.
Sourcing peptides for chromatography work in Canada
If you are buying research peptides in Canada, the instrument knowledge above becomes a few practical questions.
- Ask how purity was measured. Mode, column, gradient and wavelength should be stated or available. Our HPLC testing guide explains what a purity figure covers.
- Check identity separately. HPLC counts components; mass spectrometry confirms the main one is the right molecule.
- Account for counter-ions. Lyophilized peptides can carry counter-ions and water, so weighed mass is not necessarily peptide mass; see net peptide content, TFA and acetate.
- Store material so the result stays true. A clean chromatogram describes the powder when it was tested; the peptide storage guide covers keeping it that way.
GHRP-2A hexapeptide containing tryptophan and D-phenylalanine, residues that absorb strongly in low UVView product
GHRP-6Differs from GHRP-2 at its first two residues, a close pair for comparing retentionView product
SNAP-8Eight residues and no aromatic side chains, so its UV signal comes mainly from peptide bondsView product
LL-37A 37-residue cationic, amphipathic peptide, the type whose retention responds to ion-pairingView product
Research use only. Pure Peptides Canada products are laboratory reagents for in-vitro research, not for human or veterinary use. This article explains published research for education; it is not medical, dietary or dosing guidance.
Frequently asked questions
What is liquid chromatography?
Liquid chromatography is a separation technique in which a liquid mobile phase carries a sample through a stationary phase, usually particles packed in a column. Components that interact more strongly with the stationary phase are held longer, so each leaves at its own retention time and is recorded as a peak.
What is the difference between liquid chromatography and HPLC?
Liquid chromatography is the general technique; high-performance liquid chromatography (HPLC) is the instrumented form that pumps the mobile phase through a column of small particles under pressure. Classical column chromatography, as Tswett practised it, used an open glass tube packed with powder. HPLC gives sharper, faster and more reproducible separations.
How does HPLC work?
A pump delivers a mobile phase, often a programmed gradient of water and an organic solvent, through a packed column. An autosampler loads the sample into the flow through a switching valve, the components separate in the column by how strongly they are retained, and a detector records each one as it emerges. Software integrates the peaks into area percentages.
What are the main components of an HPLC system?
A typical HPLC system has solvent reservoirs with a degasser, a gradient pump, an autosampler, a column in a temperature-controlled oven, a detector and a data system. Each controls one variable of the separation.
What are the types of liquid chromatography?
The common modes are reversed-phase, normal-phase, hydrophilic-interaction (HILIC), ion-exchange, size-exclusion and affinity chromatography. They differ in what drives separation: hydrophobicity, polarity, hydrophilicity, charge, size or specific binding. Reversed-phase is the standard for peptide purity.
Why is reversed-phase HPLC used for peptides?
Reversed-phase HPLC can resolve peptides that differ by a single amino acid, and its retention behaviour is well understood and even predictable from sequence. An acidic ion-pairing reagent, usually trifluoroacetic acid, pairs with positively charged groups and sharpens peaks.
Research peptides with batch-listed reports
Lyophilized research peptides, priced in Canadian dollars and shipped tracked from inside Canada. Pay by credit or debit card, Interac e-Transfer or cryptocurrency at checkout.
References
- Tswett M. Adsorptionsanalyse und chromatographische Methode. Anwendung auf die Chemie des Chlorophylls. Ber Dtsch Bot Ges. 1906;24:384–393. Full text
- Strain HH, Sherma J. M. Tswett: Adsorption analysis and chromatographic methods. Application to the chemistry of chlorophylls (translation). J Chem Educ. 1967;44(4):238. DOI
- Martin AJP, Synge RLM. A new form of chromatogram employing two liquid phases. Biochem J. 1941;35(12):1358–1368. DOI
- The Nobel Prize in Chemistry 1952: Archer John Porter Martin and Richard Laurence Millington Synge. NobelPrize.org. Link
- van Deemter JJ, Zuiderweg FJ, Klinkenberg A. Longitudinal diffusion and resistance to mass transfer as causes of nonideality in chromatography. Chem Eng Sci. 1956;5(6):271–289. DOI
- Porath J, Flodin P. Gel filtration: a method for desalting and group separation. Nature. 1959;183(4676):1657–1659. DOI
- Horváth CG, Preiss BA, Lipsky SR. Fast liquid chromatography: investigation of operating parameters and the separation of nucleotides on pellicular ion exchangers. Anal Chem. 1967;39(12):1422–1428. DOI
- Cuatrecasas P, Wilchek M, Anfinsen CB. Selective enzyme purification by affinity chromatography. Proc Natl Acad Sci USA. 1968;61(2):636–643. DOI
- Guo D, Mant CT, Taneja AK, Parker JMR, Hodges RS. Prediction of peptide retention times in reversed-phase high-performance liquid chromatography I. Determination of retention coefficients of amino acid residues of model synthetic peptides. J Chromatogr. 1986;359:499–518. DOI
- Alpert AJ. Hydrophilic-interaction chromatography for the separation of peptides, nucleic acids and other polar compounds. J Chromatogr. 1990;499:177–196. DOI
- Ettre LS. Nomenclature for chromatography (IUPAC Recommendations 1993). Pure Appl Chem. 1993;65(4):819–872. DOI
- IUPAC Compendium of Chemical Terminology (Gold Book): chromatography; normal-phase chromatography; reversed-phase chromatography. DOI
- Chen Y, Mant CT, Hodges RS. Temperature selectivity effects in reversed-phase liquid chromatography due to conformation differences between helical and non-helical peptides. J Chromatogr A. 2003;1010(1):45–61. DOI
- Hartmann E, Chen Y, Mant CT, Jungbauer A, Hodges RS. Comparison of reversed-phase liquid chromatography and hydrophilic interaction/cation-exchange chromatography for the separation of amphipathic α-helical peptides with L- and D-amino acid substitutions in the hydrophilic face. J Chromatogr A. 2003;1009(1–2):61–71. DOI
- Shibue M, Mant CT, Hodges RS. Effect of anionic ion-pairing reagent hydrophobicity on selectivity of peptide separations by reversed-phase liquid chromatography. J Chromatogr A. 2005;1080(1):68–75. DOI
- Mant CT, Chen Y, Yan Z, et al. HPLC analysis and purification of peptides. Methods Mol Biol. 2007;386:3–55. DOI
- Kuipers BJH, Gruppen H. Prediction of molar extinction coefficients of proteins and peptides using UV absorption of the constituent amino acids at 214 nm to enable quantitative reverse phase high-performance liquid chromatography–mass spectrometry analysis. J Agric Food Chem. 2007;55(14):5445–5451. DOI
- United States Pharmacopeia. Notice of adoption of harmonized standard: <621> Chromatography (official 1 December 2022). Link
- European Medicines Agency. ICH Q2(R2) Validation of analytical procedures: scientific guideline (EMA/CHMP/ICH/82072/2006). Link
For laboratory research use only. Not for human or veterinary use. This article summarizes published research for educational purposes and is not medical advice.
