A purity figure on a certificate looks like a property of the peptide. It is really the output of a measurement — one column, gradient, detector and calculation applied to one sample — and it can only be as dependable as the method behind it. Analytical method validation is the discipline of showing, with data, that a method is fit for the job its numbers are used for.
Our HPLC guide explains in outline why HPLC purity depends on the method. This article goes deeper: the validation characteristics in ICH Q2(R2), the system suitability checks a run should pass on the day, and the measurement uncertainty testing laboratories are expected to evaluate, each turned into a question a buyer can ask. These are pharmaceutical and accreditation frameworks; we use them as yardsticks, not as claims about which applies to a research reagent or whether any laboratory’s methods are validated.
The short answer
- Validation is evidence of fitness for purpose. ICH Q2(R2), a pharmaceutical guideline adopted in November 2023, states that its objective is “to demonstrate that the analytical procedure is fit for the intended purpose”.
- Specificity comes first for a purity method. ICH Q2(R2) frames specificity and selectivity as “the extent to which other substances interfere with the determination of an analyte”.
- Accuracy and precision are different properties. In ICH Q2(R2), accuracy is closeness to an accepted reference value; precision is the “closeness of agreement (degree of scatter)” between repeated measurements of the same homogeneous sample.
- The quantitation limit sets a floor on what can be counted. For impurity tests, Q2(R2) says the quantitation limit should be equal to or below the reporting threshold.
- System suitability checks the run, not the method. Under the harmonized pharmacopoeial chromatography chapter, a quantitation peak’s symmetry factor must be 0.8 to 1.8 unless a monograph says otherwise.
- Every value carries uncertainty. Standards Council of Canada guidance requires an uncertainty estimate for quantitative methods, and any pass or fail against a limit depends on how that uncertainty is handled.
Analytical method validation
Validation of analytical procedures · method validation · showing a method is fit for purpose
- What it is
- A documented study showing that an analytical procedure performs well enough for its intended use
- Core guideline
- ICH Q2(R2), a pharmaceutical guideline adopted 1 November 2023; in effect at the EMA from 14 June 2024
- Companion
- ICH Q14, Analytical procedure development, which “complements ICH Q2”
- In Canada
- Health Canada’s drug-guidance table lists Q2(R2) as implemented on 17 October 2025 and Q14 on 12 January 2026
- Characteristics
- Specificity or selectivity, range, accuracy, precision, robustness
- Each run
- System suitability tests, part of the procedure itself
- Laboratory standard
- ISO/IEC 17025 requires validation of non-standard, laboratory-developed and modified methods
- What it is not
- Proof that a given number is right, or a statement about rules for research reagents
What a validated method can support
- The procedure can measure what it claims to
- Known accuracy, precision and limits over a stated range
- That a particular run performed as validated
- That the sample tested is the material in your vial
Validation covers the method; system suitability covers the run; the batch number links the sample to your vial.
Why a purity number needs analytical method validation
An HPLC purity is calculated, not observed. The harmonized pharmacopoeial chromatography chapter describes that calculation, normalization: each peak’s area as a percentage of the total area of all peaks, excluding solvent and mobile-phase peaks “and those at or below the disregard limit or reporting threshold”. Every input to that sum depends on the method: whether impurities separate from the main peak, whether small peaks are detected, whether the detector responds in proportion. Change the column, gradient or wavelength and the number can change with it.
Validation is the evidence that those inputs behave, which is how ICH Q2(R2) frames its objective: showing that a procedure is fit for its intended purpose. Eurachem’s guide for testing laboratories makes the same point from the other side: a result fit for its intended use “must be sufficiently reliable so that any decision based on it can be taken with confidence”. ICH adopted Q2(R2) on 1 November 2023, and it took effect at the European Medicines Agency on 14 June 2024 alongside its companion, ICH Q14 on analytical procedure development. Both are guidelines for medicines; we quote them because they set out publicly, and in detail, what a validated method should demonstrate.
Want to see where these questions land on an actual certificate? Our COA guide walks through what each field should show.
Read the COA guideThe validation characteristics in plain English
ICH Q2(R2) groups the method validation parameters into specificity or selectivity, range, accuracy, precision and robustness. The table adds the two checks that sit around them and says what each protects a purity figure against.
| Characteristic | What it means | What it protects a purity figure against |
|---|---|---|
| Specificity or selectivity | How far other substances interfere with measuring the analyte | An impurity hidden under the main peak and counted as product |
| Range and response | The interval with suitable accuracy, precision and response | A detector that stops responding in proportion at large or small peaks |
| Accuracy | Closeness to an accepted reference value | A systematic bias in the reported number |
| Precision | Scatter of repeated measurements: repeatability, intermediate precision, reproducibility | A number that would change on another run, day or analyst |
| Detection and quantitation limits | The lowest amount detectable, and the lowest quantifiable with suitable precision and accuracy | Small impurities that are never seen, or never counted |
| Robustness | Capacity to perform as expected in normal use, tested by deliberate variations | A method that drifts when conditions wander slightly |
| System suitability | Checks run with each analysis | A run in which the system was not performing as validated |
| Measurement uncertainty | The dispersion of values attributed to the measured quantity | False confidence in a number close to a specification limit |
Specificity: can the method see the impurities that matter?
For a purity method, specificity comes first, because an impurity that co-elutes with the main peak is simply counted as product, and no amount of precision or linearity data will reveal it. ICH Q2(R2) separates the two terms. Specificity describes “the ultimate state, measuring unequivocally a desired analyte”; selectivity is a relative term for how far analytes in a mixture “can be measured without interferences from other components with similar behaviour”.
For purity and impurity tests, the guideline says discrimination can be established “by stressing or spiking product” to achieve appropriate impurity levels and, where impurities cannot be prepared or isolated, by comparison with an orthogonal procedure. Peptides make this demanding. The US FDA describes peptide-related impurities as sequences “related to, but different from” the active ingredient through insertion, deletion or other modification, and for certain generic synthetic peptide drugs it recommends “sensitive and high resolution analytical procedures (e.g., UHPLC-HRMS)” to detect them. Close relatives of a sequence can elute near the main peak, which is why reading the chromatogram and LC-MS identity confirmation are complementary checks rather than alternatives.
Accuracy, precision and range
The definitions keep two often-blurred ideas apart. In ICH Q2(R2), accuracy expresses the closeness of agreement between a measured value and one “accepted either as a conventional true value or as an accepted reference value”, while precision expresses the “closeness of agreement (degree of scatter)” among repeated measurements of the same homogeneous sample. A method can be precise and wrong, repeating the same biased number.
Q2(R2) considers precision at three levels. Repeatability covers the same conditions over a short interval, and the guideline suggests at least nine determinations across the reportable range, or six at 100% of the test concentration. Intermediate precision adds “different days, different environmental conditions, different analysts and different equipment”; reproducibility compares laboratories. Accuracy can be shown with, for example, three concentrations in triplicate.
Range sets the boundaries. The guideline’s example reportable range for an assay runs from 80% to 120% of declared content, and for an impurity from the reporting threshold to 120% of its acceptance criterion. A purity figure leans on both: the main peak’s range and the impurities’. Where peptide content rather than purity is the question, an independent technique such as amino acid analysis gives a cross-check from a different principle.
Detection and quantitation limits
The detection limit is “the lowest amount of an analyte in a sample which can be detected but not necessarily quantitated as an exact value”; the quantitation limit is the lowest amount that “can be quantitatively determined with suitable precision and accuracy”. Laboratories often abbreviate them LOD and LOQ. Q2(R2) offers several ways to establish them, including signal-to-noise ratios of 3:1 for detection and at least 10:1 for quantitation, or calculation from the standard deviation of the response and the calibration slope (3.3σ/S and 10σ/S).
For a purity figure, the quantitation limit matters more than it looks. Q2(R2) says that for impurity tests it “should be equal to or below” the reporting threshold, and the harmonized chromatography chapter writes the same requirement into system suitability. Because normalization leaves out peaks at or below the reporting threshold, a method that cannot see small impurities does not report them as impurities: it reports a higher purity. For related-substances tests in pharmacopoeial monographs, the chapter notes, the reporting threshold “is generally 0.05%”.
System suitability: checking the system on the day
Validation happens before a method enters routine use; system suitability is checked every time it runs. Q2(R2) calls the system suitability test “an integral part of analytical procedures”, and its glossary, in wording taken from ICH Q14, says such tests verify that the measurement system and associated operations “are fit for the intended purpose and increase the detectability of unacceptable performance”.
For chromatography, the chapter harmonized across the European, Japanese and United States pharmacopoeias (Ph. Eur. 2.2.46, JP 2.00, USP <621>) puts numbers on this. System sensitivity is defined by signal-to-noise; the quantitation peak’s symmetry factor must be 0.8 to 1.8 unless a monograph says otherwise; resolution or peak-to-valley ratio may be specified; and for assays, unless a monograph sets its own requirement, the relative standard deviation of replicate sample injections of a reference solution has a ceiling tied to the upper content limit. For an upper limit of 102.0% and six replicates, that ceiling is 0.85%. The revised chapter became official in the USP on 1 December 2022 and took effect in the Ph. Eur. on 1 January 2023, and the EDQM notes that compliance with the system suitability test is always required when chromatographic conditions are adjusted.
Measurement uncertainty and pass/fail decisions
The international metrology vocabulary defines measurement uncertainty as a “non-negative parameter characterizing the dispersion of the quantity values being attributed to a measurand”. It is usually reported as an expanded uncertainty, the standard uncertainty multiplied by a coverage factor k; NIST notes that when the normal distribution applies, k = 2 “defines an interval having a level of confidence of approximately 95 %”. A purity of 98.6% ± 0.5% (k = 2), to take a hypothetical figure, says far more than a bare 98.6%.
Standards Council of Canada guidance says a robust estimation of measurement uncertainty “must be performed as part of the validation”, and Eurachem observes that most of the information needed can be obtained during validation. Reporting it is conditional: ISO/IEC 17025:2017 calls for uncertainty in a report when it is relevant to the validity or application of the result, when a customer requires it, or when it “affects conformity to a specification limit”. Under simple acceptance, a value exactly at the limit carries, in UKAS’s words, “an associated risk of false decision of 50 %”; a guard band, derived from the uncertainty, exists “with the purpose of limiting the risk of false acceptance”.
Questions to ask about a certificate’s method
None of this asks a buyer to repeat the validation; it lists questions a competent laboratory, or a supplier relaying its answers, should be able to handle.
- Which procedure produced the number? Technique, column, gradient and detection wavelength, or a method reference.
- Was it validated, or verified, for this peptide? SCC guidance applies validation to non-standard, laboratory-developed and modified methods, and verification to standard methods adopted without significant change.
- How was specificity shown? Which impurities were spiked or generated by stress, and was an orthogonal technique such as LC-MS used?
- What were the quantitation limit and reporting threshold? Peaks below them are not in the purity figure.
- Did the run meet system suitability, and against which criteria?
- What is the uncertainty, and which decision rule stands behind any “pass”?
- Does the report tie the tested sample to your batch? Our guide to what a certificate should show covers that check.
A certificate need not answer all seven on its face. What matters is whether the answers exist when asked.
What’s established and what isn’t
Established, and citable
- ICH adopted Q2(R2) on 1 November 2023; it took effect at the EMA on 14 June 2024 (ICH; EMA).
- Q2(R2) defines specificity, accuracy, precision and the detection and quantitation limits (ICH Q2(R2)).
- For impurity tests, the quantitation limit should be at or below the reporting threshold (ICH Q2(R2)).
- System suitability is integral to chromatography; the symmetry factor is 0.8–1.8 unless stated otherwise (harmonized chapter).
- ISO/IEC 17025 requires validation of non-standard, laboratory-developed and modified methods (SCC RG-MVVT).
- Simple acceptance carries a 50% risk of false decision for values at the limit (UKAS LAB 48).
Open, or outside what validation can show
- Which of these frameworks, if any, applies to a research reagent. This article does not address it.
- Whether a given laboratory’s method is validated, or a given run passed system suitability; only documentation can show it.
- Whether validation covered every impurity a synthesis produced; specificity is shown only for the impurities tested.
- How two laboratories’ purity figures compare when their methods differ.
- Whether a stated uncertainty captures every significant source; SCC notes that validation studies may miss some.
Applying this when comparing suppliers in Canada
The same vocabulary now sits in Canadian drug guidance: Health Canada’s table of implemented ICH guidelines lists Q2(R2) from 17 October 2025 and Q14 from 12 January 2026. That is pharmaceutical guidance, and we make no claim that it applies to research reagents; it is useful here as a current, shared definition of what “validated” should mean.
For testing laboratories, the Standards Council of Canada publishes requirements and guidance on method verification and validation, describing validation as ensuring a method has “the capabilities and performance characteristics consistent with specified requirements” and is fit for its intended use. If a report names an accreditation, it is reasonable to ask which methods that accreditation covers.
In practice, compare suppliers on whether these questions get answered, not on the size of the headline number. We are a supplier too, so put the same questions to us: the certificates we publish are on our Lab Testing page, and method questions can go through the contact button below.
Read the certificate behind the number
Janoshik test reports for the research peptides below are listed on our Lab Testing page, so you can read each purity result in context rather than as a bare number on a label.
Research use only. Pure Peptides Canada products are laboratory reagents for in-vitro research, not for human or veterinary use. This article explains analytical method validation, system suitability and measurement uncertainty for education; it is not medical, dietary or dosing guidance, and it is not legal or regulatory advice.
Frequently asked questions
What is analytical method validation?
Analytical method validation is the documented demonstration that a measurement procedure performs well enough for its intended use. ICH Q2(R2), a pharmaceutical guideline, frames the objective as showing that a procedure is fit for the intended purpose, through characteristics such as specificity, range, accuracy, precision and detection limits. Laboratories accredited to ISO/IEC 17025 must validate non-standard, laboratory-developed and modified methods.
What does ICH Q2(R2) cover?
ICH Q2(R2) covers the validation of analytical procedures used for release and stability testing of commercial drug substances and products, and other procedures on a risk basis. It defines the validation characteristics and suggests study designs for each. ICH adopted it on 1 November 2023, and Health Canada lists it as implemented from 17 October 2025.
What is the difference between accuracy and precision?
Accuracy is how close a measured value is to an accepted reference value; precision is how closely repeated measurements of the same homogeneous sample agree with one another. A method can be precise but inaccurate, repeating the same biased value. ICH Q2(R2) considers precision at three levels: repeatability, intermediate precision across days, analysts and equipment, and reproducibility between laboratories.
What are LOD and LOQ?
LOD and LOQ are the limit of detection and the limit of quantitation, the detection and quantitation limits of ICH Q2(R2). The detection limit is the lowest amount that can be detected but not necessarily quantified; the quantitation limit is the lowest amount that can be measured with suitable precision and accuracy. ICH Q2(R2) accepts signal-to-noise ratios of 3:1 and 10:1 as one way of estimating them.
What is system suitability in HPLC?
System suitability is a set of checks run with each analysis to confirm that the chromatographic system is performing adequately that day. Under the harmonized pharmacopoeial chapter they can include signal-to-noise, a symmetry factor of 0.8 to 1.8 for the quantitation peak, resolution and the repeatability of replicate sample injections. Validation shows a method can work; system suitability shows a given run did.
Does a validated method mean the purity result is correct?
Not on its own. Validation shows that a procedure is capable of accurate, precise and specific measurement over a stated range. Whether a particular number is right also depends on the run meeting system suitability, the sample being the material described, and the uncertainty of the value. A validated method can still miss an impurity it was never challenged with.
Ask how the number was measured
See the certificates on our Lab Testing page, or send us a question about a method.
References
- International Council for Harmonisation. Q2(R2): Validation of analytical procedures. Final version, adopted 1 November 2023. ICH
- European Medicines Agency. ICH Q2(R2) Guideline on validation of analytical procedures, Step 5 (EMA/CHMP/ICH/82072/2006). Date for coming into effect 14 June 2024. EMA
- European Medicines Agency. ICH Q14 Guideline on analytical procedure development, Step 5 (EMA/CHMP/ICH/195040/2022). Date for coming into effect 14 June 2024. EMA
- Health Canada. International Council for Harmonisation (ICH): Quality guidelines (ICH guidelines implemented by Health Canada). Date modified 15 May 2026. Government of Canada
- Japanese Pharmacopoeia. General Tests 2.00 Chromatography (harmonized with the European Pharmacopoeia and the United States Pharmacopeia). PMDA
- United States Pharmacopeia. Notice of adoption of harmonized standard: <621> Chromatography. Posted 19 November 2021; official 1 December 2022. USP
- European Directorate for the Quality of Medicines & HealthCare. Ph. Eur. Commission adopts harmonised general chapter 2.2.46. Chromatographic separation techniques. 15 February 2022. EDQM
- US Food and Drug Administration. ANDAs for certain highly purified synthetic peptide drug products that refer to listed drugs of rDNA origin: guidance for industry. May 2021. FDA
- Standards Council of Canada. SCC requirements and guidance for method verification and validation in testing laboratories (RG-MVVT). 24 May 2021. SCC
- International Laboratory Accreditation Cooperation. ILAC G17:01/2021 ILAC guidelines for measurement uncertainty in testing. UKAS-hosted copy
- United Kingdom Accreditation Service. LAB 48: Decision rules and statements of conformity, edition 5. July 2024. UKAS
- Joint Committee for Guides in Metrology. JCGM 200:2012 International vocabulary of metrology (VIM), entry 2.26, measurement uncertainty. BIPM
- National Institute of Standards and Technology. Expanded uncertainty and coverage factors. NIST Reference on Constants, Units, and Uncertainty. NIST
- Cantwell H (ed.). Eurachem Guide: The fitness for purpose of analytical methods: a laboratory guide to method validation and related topics. 3rd edition, 2025. Eurachem
For laboratory research use only. Not for human or veterinary use. This article summarizes published standards and regulatory material for educational purposes and is not medical, legal or regulatory advice.